Recently rules for auto-repeat were created, but I saw that in all cases the driver (ASTM protocol) is sending the order (action code) as new/adding tests in IDM/Prime.

Exemple of message sent to instrument (IDM/Prime):

Host -> : H|\^&|||LIS^1.00^1.00|||||||P|1|20220310075448[CR]

Host -> : P|1|123456|||PATIENT NAME||19720101|F[CR]

Host -> : O|1|2206907085D^^^||^^^a-IgE^tIgE^1^1^|||20220310075416||||N[CR]

Host -> : L|1|F[CR]

In segment O (pipe 12) always is sendind "N" and in manual is describe action code "N" for "New test request accompanying a new specimen. If the specimen already exists, the test is added. If the test already exists, it is ignored as a duplicate transmission."
In repetition cases test already exists then it is ignored.