Adapters Request Portal

VITROS5k6 - Reagent informations

The instrument sends to CommServer informations on reagent used to analyse the test (in the extended message record M).


Exemple:


R|1|^^^1.0000+301+1.0|8.1|g/dL||^0^VS\^0^\^0^\^0^||V|||20220329171709|20220329172330|76001093

M|1|X|06226160^20220425153110^20220329154659^F9352^D9233|20220328162313^N^20241221162313|||N


Those informations are not used by CommServer and loose in the xml trace sent by CommServer to the middleware.


<Sample Control="false" SampleID="CHIMI1" InstrumentSpecimen="6">

<Result TestCode="301" InstrumentDilution="1.0">

<ResultPart Part="1" Unit="g/dL">8.1</ResultPart>

<Flag Value="VS" />

</Result>

  • Guest
  • Apr 8 2022
  • Released
  • Attach files