8800 triple quadrupole ICP-MS (Inductively Coupled Plasma – Mass Spectrometer)

  1. instrument with model info

  • Instrument: 8800 triple quadrupole ICP-MS (Inductively Coupled Plasma – Mass Spectrometer)

  • Model: G3663A

  • Software: MassHunter

https://www.agilent.com/en/product/atomic-spectroscopy/inductively-coupled-plasma-mass-spectrometry-icp-ms/icp-ms-software/icp-ms-masshunter-software

2-manufacturer

  • Agilent

3-integrations points, (what info needs to be exchanged ) ie: What information do they need to transfer? Peaks? Absorbance?

  • Concentration of trace elements (e.g. copper, zinc, selenium, etc which are test names)

  • Full list of elements/test names can be provided (please let me know)

4-use of the instruments: is the instrument used for diagnostic or research labs

  • The instrument is not Health Canada approved and primarily for research use, but they are used by clinical diagnostic labs with lab developed tests (LDTs).


The software has LIMS feature that exports data as .csv file. There is flexibility in what that report contains, and column headings have predefined names (please let me know if you need anything else). We can define which columns are included in report and we have flexibility to export only patient results (i.e. no calibrators, QC, etc) if it will help.


The data from .csv file that would be used to export data to Delphic through Infinity is following:

  • Sample Name: contains sample ID (scanned with barcode from sample tube)

  • Element Full Name: will be used to match the order (i.e. assay name)

  • Concentration: result to be reported

I am not sure if any other information is required to gather data from the software and then move it to Delphic. Typically patient report will include: (i) element name (can be multiple ordered on the same specimen) and (ii) measured concentration.


  • Guest
  • Jan 12 2023
  • More info required
  • Attach files
  • Guest commented
    January 25, 2023 19:08

    Hello Poala,
    here is the details provided by the vendor ,

    Hi Mina and Robert,

    Thank you for your time yesterday.

    Please see attached LIMS files documents and sequence templates info as
    discussed. Please let us know if this is what you are looking for.

    I also wanted to come back on something that stuck with me and I believe
    there is a risk involves. You mention that you wanted the users to see the
    final report on the middleware software and then send for approval, but
    this is unusual in what I have seen in clinical spaces. The GCMSMS is not
    like a pH meter or a minerals analysis in plasma where the final number is
    fixed and matrix complexity will not interfere. With a GCTQ, there are some
    sample result evaluation and peak integration required by the biochemist
    prior to result approval due to the matrix complexity and analysis
    complexity. What I would suggest is to have the biochemist perform the
    results authorization on Agilent SW and then when they are satisfied, they
    can send it to the middleware that will send it to the LIMS���

    If you want to have access to the results from anywhere in the hospital,
    then they can put the computer in a server environment or connect to the
    workstation using remote desktop connections.

    This could be discussed further at our next TC with your development team
    and Shared Health IT and laboratory department.

    Regards,

    *Philippe Lebel, M.Sc.*

    Mass Spectrometry Product Specialist

    Life Science and Applied Markets

    Agilent Technologies, Inc.

    T: 514-832-4841 |

    philippe.lebel@agilent.com

    --------

    *Mina Mikail*
    Marketing Specialist - IT IS
    Roche Core Lab & Digital Solutions/ Solutions Roche Laboratoires centraux

    [image: Learn More]
    Roche Diagnostics,
    Division de/of Hoffmann-La Roche Limit��e/Limited
    201, Boul. Armand-Frappier
    Laval, Qu��bec, H7V 4A2, Canada

    Tel: +1 514 458-7921
    Email: Mina.Mikail@roche.com
    www.rochecanada.com

    *Confidentiality Note*: This message is intended only for the use of the
    named recipient(s) and may contain confidential and/or proprietary
    information. If you are not the intended recipient, please contact the
    sender and delete this message. Any unauthorized use of the information
    contained in this message is prohibited.

    *Note de confidentialit�� :* Ce message est destin�� uniquement �� l'usage
    du/des destinataire (s) et peut contenir des informations confidentielles
    et/ou privil��gi��es. Si vous n'��tes pas le destinataire, veuillez s.v.p.
    communiquer avec l'exp��diteur et supprimer ce message. Toute utilisation
    non autoris��e des informations contenues dans ce message est interdite.

    Si vous d��sirez obtenir des informations sur les produits et services de
    Roche Diagnostics, cliquez sur M���INSCRIRE
    *.* Si vous recevez un
    message ��lectronique commercial et que vous ne souhaitez plus en recevoir,
    cliquez sur ME D��SABONNER
    .

    Should you wish to obtain more information about Roche Diagnostics products
    and services, click on SUBSCRIBE
    . If you receive a
    commercial electronic message from Roche Diagnostics and you no longer wish
    to receive such messages in the future, click on UNSUBSCRIBE
    .

  • Paola Maita Gonzalez commented
    January 13, 2023 14:08

    Hello Mina,

    Thanks for contacting us.

    After reviewing your request, we have detected that the document attached is not a host interface manual.

    For being able to check if any of our available drivers is what you need, it's necessary to have the instrument interface manual of the instrument, which contains the message description sent from and to the instrument.

    Do you think you can provide this document via email before end of January?

    Thanks,

    Paola