Our customer is planning to connect analyzer: UriSed 3 PRO (77 ELEKTRONIKA) to Cobas Infinity. Does driver "647 - 1 - UriSed v1.0.2_NA_en" is compatible and can be used with UriSed 3 PRO?
Message examples attached to this idea post as a separate PDF document called "UriSed_Communication_Examples_v1 (1) (1).pdf". Regarding protocol version, which to use - manufacturer info is that it has to be used - "Bidirectional from 1.7", so examples also should be checked for this protocol version, please see in communication example file pages 11-13. Please let me know if any more info would be required from our side.
Is it possible for you to get us real message traces for patient, query and QC results from Labumat? Considering this instrument has its own protocol, it would be the best way to ensure an optimal result.
Also, Can you please confirm which of the proprietary protocols use the instrument you're connecting? In the manual we see these two: Bidirectional up to 1.6.1 and Bidirectional up to 1.7. Without this information is not possible to continue with the development.
Hello Rolands,
Can you please share the PDf via email? Unfortunately, I didn't receive it here in AHA!
Thanks for confirming that the Bidirectional up to 1.7 protocol is the one needed.
Regards,
Paola
Hello Paola,
Message examples attached to this idea post as a separate PDF document called "UriSed_Communication_Examples_v1 (1) (1).pdf".
Regarding protocol version, which to use - manufacturer info is that it has to be used - "Bidirectional from 1.7", so examples also should be checked for this protocol version, please see in communication example file pages 11-13.
Please let me know if any more info would be required from our side.
Thanks,
Rolands
Hello Rolands,
Thanks for your request.
Is it possible for you to get us real message traces for patient, query and QC results from Labumat? Considering this instrument has its own protocol, it would be the best way to ensure an optimal result.
Also, Can you please confirm which of the proprietary protocols use the instrument you're connecting? In the manual we see these two: Bidirectional up to 1.6.1 and Bidirectional up to 1.7. Without this information is not possible to continue with the development.
Thanks,
Paola