Adapters Request Portal

New feature : T511

Rerun Reflex

  • Nicolas Gagnaire
  • Apr 18 2023
  • Will not implement
  • Attach files
  • Paola Maita Gonzalez commented
    April 25, 2023 07:02

    Hello Nicholas,

    Thanks for providing the instrument manual. After reviewing it, we have seen that this feature is described as an internal feature in the instrument. The rerun information going from the instrument to the LIS acts like a flag, not as an action code. It's sent to the middleware but then the instrument ignores it when the middleware sends it back. Therefore, no rerun/reflex action can be triggered by the middlewares.

    In summary, this request cannot be implemented as rerun/reflex action codes sent from the middleware.

    Thanks,

    Paola

  • Paola Maita Gonzalez commented
    April 19, 2023 09:42

    Hello Nicholas,

    Thanks for contacting us.

    The manual attached to the request is for cobas t 711 instrument. Can you please confirm if it's the manual you need us to support for cobas t 511?

    Thanks,

    Paola