Adapters Request Portal

update driver DxH500 (QC produc mapping include control name and control level)

update driver DxH500 (QC product mapping include control name and control level). current: driver DxH only capture control name to QC product mapping. customer want have option : QC product mapping include control name and control level.

example:

H|\!~|||DxH|||||LIS||Q|LIS2-A|20080925122107

P|1|||

O|1|130000001!Level 2!COULTER® 6C Cell!20080813|00001||||||||Q|||||||||OL3!OL3!12345-1-

B1|!M!E!Shift 1||||F

C|1|I|System Event: RBC|I!Y

C|2|I|System Event: PLT|I!Y

C|3|I|System Event: WBC|I!Y

C|4|I|Data Disc: D|I!Y

R|1|!!!WBC!33256-9|5.3!R L |10^3/uL|||A||F||||20080925093226|IEP005_DxH800

R|2|!!!RBC!789-8|5.37!R H |10^6/uL|||A||F||||20080925093226|IEP005_DxH800

R|3|!!!HGB!718-7|13.7! H |g/dL|||A||F||||20080925093226|IEP005_DxH800

R|4|!!!HCT!4544-3|42.9!R H |%|||A||F||||20080925093226|IEP005_DxH800

R|5|!!!MCV!787-2|79.9!R L |fL|||A||F||||20080925093226|IEP005_DxH800

R|6|!!!MCH!785-6|25.5!R L |pg|||A||F||||20080925093226|IEP005_DxH800

R|7|!!!MCHC!786-4|31.9!R L |g/dL|||A||F||||20080925093226|IEP005_DxH800


QC producmapping = COULTER® 6C + Level 2


current: QC producmapping = COULTER® 6C

  • Guest
  • Jun 12 2023
  • Released
  • Attach files
  • Paola Maita Gonzalez commented
    June 12, 2023 09:48

    Hello Binh,

    According to the manual, I see that:

    • Level 2 = Control level

    • COULTER® 6C Cell = Control type / control name



    Does this mean that you need to concatenate the Control level and Control type?