Thanks for reaching out. As you may see in the roadmap, this driver delivery is currently planned for mid-October. If there are any changes in the delivery dates, it'll be updated there.
We contected with Analyzer´s Vendor and we got a confirmation, this analyzer will not need to receive a order with dilution, cos all assays uses ELISA methodology.
On a first evaluation, we have seen that in the manual attached, for ASTM protocol, the instrument sends patient results and query messages only. Therefore, the rest of the information you requested to be supported is not possible to be supported by the driver.
On the same line, is it possible that you to share with us instrument real traces? It's always nice to compare the instrument real behavior with the manual description.
Especially we need to see the flag list and their description. We see inconsistencies between the defintion and the examples. In the definition No liquid detected, and then in the example messages we see No liquid. Which would be the one sent by the instrument?
We also need the dilution list, if you need us to implement it. The information needed would be like 1:1 and its meaning.
Hello Cristiano,
Thanks for reaching out. As you may see in the roadmap, this driver delivery is currently planned for mid-October. If there are any changes in the delivery dates, it'll be updated there.
Regards,
Paola
Hi Paola,
G. Afternoon,
Just to FUP, for the next steps is it possible to have estimated date for release?
If you need more information I will at your disposal.
Regarding,
Cristiano
Hello Cristiano,
Thanks for the update!
Regards,
Paola
Hi Paola,
G. Afternoon,
We contected with Analyzer´s Vendor and we got a confirmation, this analyzer will not need to receive a order with dilution, cos all assays uses ELISA methodology.
Regarinding error flags there are only three:
"No Liquid" , "No Liquid Detected" and "MAN ID"
Regarding,
Cristiano
Hello Cristiano,
Thanks! Please remember we also need to see all the flags.
Regards,
Paola
Hi Paola,
We will try to extract those example of Dilution next monday, one of RPS will be on site to see some workaroudn to generate those traces.
Hello Cristiano,
Thanks for this example.
Can you please confirm the rest of the flags? It'd be nice to confirm the complete flag list.
About the dilution, I guess that the right way to go is to not support it then. Is that ok?
Thanks,
Paola
Hi Paola,
G. Afternoon,
As requested, our customer generated few examples of Query, Result and No Liquid Flags.
Regarding dilution, this customer do not have test that use Dilution, so they dont have ways to generate an example.
See attached File, let me know if you need more infos or examples.
Hi Paola,
G. Morning,
Am so sorry for huge delay, I didnt noticed AHA e-mail regarding this post.
I will FUP with Analyzer Vendor to see if there is some trace files that contain example of Query, Results, Flags(No Liquid) and Dilution.
Thanks in Advanced
Hello Cristiano,
Thanks for reaching out.
On a first evaluation, we have seen that in the manual attached, for ASTM protocol, the instrument sends patient results and query messages only. Therefore, the rest of the information you requested to be supported is not possible to be supported by the driver.
On the same line, is it possible that you to share with us instrument real traces? It's always nice to compare the instrument real behavior with the manual description.
Especially we need to see the flag list and their description. We see inconsistencies between the defintion and the examples. In the definition No liquid detected, and then in the example messages we see No liquid. Which would be the one sent by the instrument?
We also need the dilution list, if you need us to implement it. The information needed would be like 1:1 and its meaning.
Regards,
Paola