request Driver for instrument Monitor (replace driver Monitor-S.wco)

driver "166 - 1 - Monitor-S - v1.0.1_NA_en" only capture result 1h of erythrocyte sedimentation. we have tried connect instrument with cobas infinity (driver 166 - 1 - Monitor-S - v1.0.1_NA_en). when instrument send resul to cobas infinity have include result erythrocyte sedimentation 1h, 2h, ik (h), but cobas infinity only capture result 1h, don't capture result 2h, ik (although we have config at test mapping)

  • Guest
  • Jul 8 2023
  • Released
  • Attach files
  • Paola Maita Gonzalez commented
    July 12, 2023 15:28

    Thanks, Binh!

  • Guest commented
    July 12, 2023 14:13

    Hi Paola

    current, vendor share for me user manual and log from instrument

    Thank so much Paola

  • Paola Maita Gonzalez commented
    July 12, 2023 13:13

    Hello Binh,

    I have been searching in the archives for the manual used for this development, and I haven't found any. Is there any chance that you may contact the vendor for this?

    Thanks,

    Paola