Sysmex SP-50

Our driver for the Sysmex slide maker SP-10 doesn't support the SP-50.

The interface is very similar.

  • New HIM v2.0 added (02/Sep/206)

New feature of the SP-50:

The middleware can tell the SP-50 to rinse multiple time after aspiration of the sample. This is recommended for samples with a high white blood cell count (WBC). It reduces the carry-over risk for the next sample.

  • The new HIM explains that the host can send 0, 1, 2 or 3 as values for rinsing (see screenshot and HIM p23 & p67).

  • Important: This value is decided based on the WBC value of the current sample. The user should be able to configure a threshold. e.g. If WBC>30 then Rinse = 2.

  • The default is 0 and should be sent if no threshold was configured or conditions don't match.

  • We are already able to map the WBC value because it's sent in the query reply to the SP.

  • Guest
  • Jul 21 2023
  • More info required