Atellica driver not working

We have an emergency on the Atellica coag driver.

The patient grows impatient.

Siemens helps us solve the problem.

We held a meeting and the supplier identified points to be corrected.

The analyser stop to send query after the answer to first one. Siemens says this following points are important and may be the origin of the issue

1/ order -> 9.4.26 only use Q

2/ order -> 9.4.6 delete ^^

3/ Don't send message without P and O trames

For information : Atellica driver is very close to BCSXP driver

  • Guest
  • Feb 24 2022
  • Released
  • Attach files
  • Guest commented
    February 24, 2022 14:27

    2022-02-16 13:35:04,832

    H|\^&|||ACOAG360^1.0^10170^H1P1O1R1C1Q1L1^1.0|||||||P|1|20220216133940

    Q|1|^220451095014||^^^ALL||||||||O

    L|1


    2022-02-16 13:35:04,998

    [ENC]<ENQ>

    2022-02-16 13:35:05,060

    [ENC]<STX>1H|\^&|||system_LIS|||||LIS_ID||P|1

    <ETX>7A

    <LF>

    2022-02-16 13:35:05,138

    [ENC]<STX>2P|1|808567|||SILVA SCHULZE^ESTEBAN||20220209|M

    <ETX>E0

    <LF>

    2022-02-16 13:35:05,216

    [ENC]<STX>3O|1||220451095014^^|^^^11200\^^^11210|R|20220216122905|||||||||||||||||||O\Q

    <ETX>85

    <LF>

    2022-02-16 13:35:05,294

    [ENC]<STX>4L|1|F

    <ETX>FF

    <LF>

    2022-02-16 13:35:05,373

    [ENC]<EOT>