Currently, the driver does not support QC recording.
The MAGLUMI instrument driver release notes include:
"N/A, the instrument does not send different types of results. Quality Controls. Quality controls are not provided."
However the instrument sends a control result message as LIAISON instrument.
Example:
SAMPLEID |
DESCRIPTION |
123456 |
Patient results |
#123456# |
QC results |
Maglumi trace (control quality results)
H|\^&||PSWD|Maglumi User|||||Lis||P|E1394-97|20220621
P|1
O|1|#069210212#||^^^GAD65
R|1|^^^GAD65|32.41|IU/mL|1 to 17|H||||||20220621115704
L|1|N
Hi,
QC Result :
H|\^&||PSWD|Maglumi 2|||||Lis||P|E1394-97|20200408
P|1
O|1|#27220002021#||^^^2019-nCoV IgG
R|1|^^^2019-nCoV IgG|0|AU/mL|0 to 1|N||||||20200407191034
L|1|N
regards
Hi Paola,
We have some Maglumi connected to MPL in France, I search for some documentation about QC implementation, and I will give it to you, because we are also interesting by QC in France.
BR
Damien
Hello all,
Thanks for all your votes.
We appreciate Lucas' effort in getting in touch with the instrument engineering team. Although they confirm this behavior, we would need this to be properly documented in the instrument's host interface manual.
Unfortunately, we are not able to implement behaviors that are not documented in the instrument's manual, especially if it implies delivering patient or QC results.
If anyone finds a host interface manual which documents this behavior, please reach us.
Regards,
Paola