QC - Maglumi

Currently, the driver does not support QC recording.

The MAGLUMI instrument driver release notes include:

"N/A, the instrument does not send different types of results. Quality Controls. Quality controls are not provided."

However the instrument sends a control result message as LIAISON instrument.

Example:

SAMPLEID

DESCRIPTION

123456

Patient results

#123456#

QC results



Maglumi trace (control quality results)

H|\^&||PSWD|Maglumi User|||||Lis||P|E1394-97|20220621

P|1

O|1|#069210212#||^^^GAD65

R|1|^^^GAD65|32.41|IU/mL|1 to 17|H||||||20220621115704

L|1|N

  • Guest
  • Jul 12 2022
  • Will not implement
  • Attach files
  • Guest commented
    September 01, 2022 09:28

    Hi,


    QC Result :

    H|\^&||PSWD|Maglumi 2|||||Lis||P|E1394-97|20200408

    P|1

    O|1|#27220002021#||^^^2019-nCoV IgG

    R|1|^^^2019-nCoV IgG|0|AU/mL|0 to 1|N||||||20200407191034

    L|1|N


    regards

  • Guest commented
    September 01, 2022 08:45

    Hi Paola,


    We have some Maglumi connected to MPL in France, I search for some documentation about QC implementation, and I will give it to you, because we are also interesting by QC in France.


    BR


    Damien

  • Paola Maita Gonzalez commented
    August 03, 2022 07:07

    Hello all,


    Thanks for all your votes.


    We appreciate Lucas' effort in getting in touch with the instrument engineering team. Although they confirm this behavior, we would need this to be properly documented in the instrument's host interface manual.


    Unfortunately, we are not able to implement behaviors that are not documented in the instrument's manual, especially if it implies delivering patient or QC results.


    If anyone finds a host interface manual which documents this behavior, please reach us.


    Regards,

    Paola

  • Guest commented
    July 12, 2022 20:30

    voted