current, enginer of siemens have enable option at CU of instrument :"Send reagent information to lis"
so, cobas infinity was recived H|\^&|||UIW_LIS|||||LIS_ID||P||20220815060637 P|1 O|1|BIORAD1^011218^A10||^^^Trig^^^1|R|20220815060000|||||Q||||Serum||||||||||F M|1|SHD^CEN:NG^V1^O|BIORAD1|26491 R|1|^^^Trig^^^1^RLU^1857164#0|184.000759413459|||||F||hmuh^No Review||20220815060636|SP00686^CM00785 C|1||PREAG^211622;21162210202022110102231;2|G C|2||CALINFO^Lot=326084A;DateTime=7/6/2022 8:56:48 AM;Status=Valid;ID=2394;|G R|2|^^^Trig^^^1^DOSE^1857164#0|2.15|mmol/L|1.93 - 2.29|||F||hmuh^No Review||20220815060636|SP00686^CM00785 C|1||PREAG^211622;21162210202022110102231;2|G C|2||CALINFO^Lot=326084A;DateTime=7/6/2022 8:56:48 AM;Status=Valid;ID=2394;|G L|1|N
current: 211622;21162210202022110102231;2 --> lot ID ( driver atellica 3.2.0.24) we want : 211622 --> lot ID, 21162210202022110102231 --> reagent ID