Field for "Instr. sample type" should be mandatory with driver versions higher than cobas t711 3.7.0.51

With the new driver version cobas t711 3.7.0.51 the test ordering was not possible with infinity.


After a long troubleshooting time interval, we came to the conclusion that the problem is the missing sample type definition with the abbreviation PLAS.


The box for "Instr. sample type" is not mandatory to fill in something. But without the information the instrument does nothing.


A senseful conclusion would be to set this field to mandatory in the driver versions.