The instrument ILACLTOP is able to send the rack number and the tube position in the rack. But Infinity doesn't retrieve it and our customer would like to have this information sent from Infinity to the LIS.
Please find below, in red the rack number and tube position in the rack from ACLTOP's message
H|@^\|12931706A0134EA88FF1FDEB4489FEF9||ILACLTOP_1|||||INFINITY||P|1394-97|20230412155050
P|1||7158321||TEST^PATIENT||19011231|M|||||^
O|1|023061479401|<1681306012_6382>|^^^921|R|20230412152652|||||||||P||||||||||O@F
R|1|^^^921|31.2|s||N||F@V||SysAdmin^SysAdmin||20230412153306|ILACLTOP_1^06^1
C|1|I|5770^H�moglobine Haute^PW|I
C|2|I|LI-TCA SYN^N0229774^2024-2|I
C|3|I|LI-TCA SYN CaCl2^N0229937^2025-2|I
C|4|I|LI-NORMAL^N0723952^2025-7|I
C|5|I|LI-ANORMAL^N0724169^2025-7|I
R|2|^^^921|1.06|Ratio||N||F@V||SysAdmin^SysAdmin||20230412153306|ILACLTOP_1^06^1
C|1|I|5770^H�moglobine Haute^PW|I
L|1|N
Please let me know in case you need more information
Kind regards