The latest CS driver for the analyzer: Biorad D10.
System: Infinity ver. 3.3.14
Problem:
The issue relates to the incorrect assignment of test results for patient orders and for QC orders. Each barcode, which according to Infinity's configuration is TubeID containing a suffix, is treated by the driver as a control, e.g. 10000014-2 (attached trace named "1 PROBLEM_ID8znakow_z_myslnikiem", incorrect assignment of order ID 10000014-2 as a control in the attachment "Biorad10 - kontrole"). So the result of a specific patient's test is added in QC, not to this order.
If the barcode of the material has no suffix, e.g.: 1000000019 - then the result is treated correctly (as the result of the order) (attached trace named: "2 PROBLEM_ID10znakow - OK"). I also attach a third trace file called "3 PROBLEM_KONTROLA" - this is a trace containing a real QC control.
Summing up: All result received from Biorad D10 with alphanumeric sample id's are considered as QC result.